Zen-O lite is manufactured to meet the exacting standards of the European Medical Device Directive. GCE Healthcare ensures
that our products undergo rigorous safety and quality checks before they reach our customers. Below is the technical information
for the Zen-O lite??portable oxygen concentrator.
Size:
??249 mm (W), 97 mm (D), 235 mm (H)
??9.8″ (W), 3.8″ (D), 9.25″ (H)
??Unit weight: 2.5 kg (5.5 lbs) (without carry bag)
Power requirements:
??AC adaptor: 100-240V AC(+/- 10%)
??50-60 Hz in, 24V DC, 5.0A out
??DC adaptor: 11.5 – 16V DC in, 24V, 5.0A out
??(Important: see accessories list for model and part numbers for AC and DC power supply packs.)
??Purity: 87% – 96% at all flow rates, over operating conditions
??Setting: User adjustable in 0.5 increments from 1.0 to 5.0
??Inspiratory trigger sensitivity: -0.12cm/H20
??Setting indicator: LCD
??Maximum oxygen discharge pressure: 20.5 psi
??Humidity range: 5% to 93% ? 2% non-condensing
??Disconnect Device: Power supply input connector
??Operating altitude: 0m to 4,000m (0? to 13,000?) relative to sea level, 1060 down to 575 mbar
Sound level:
??Sound Pressure level of 37dBA at setting 2 and 18 BPM: (when measured at 1 m from front of device)
??Sound Pressure level of 48.7 dBA/Sound Power level of 56.7 dBA at setting 5
??Type of protection (electrical): Class II
??Degree of protection (electrical): Type BF
??Degree of protection (water):
??IP22 in carry bag (protection against small objects and tilted dripping water)
??IP20 out of carry bag (protection against small objects and no protection agains water entering the concentrator)
Degree of safety (flammable anaesthetic mixture):
??Not suitable for use in the presence of a flammable anaesthetic mixture
??Operating temperature:
??Continuous operating at temperatures between 5?C (41?F) and 40?C (104?F).
Storage temperature:
??Between -20?C (-4?F) and 60?C (140?F).
??Alarm sound pressure range: 68.1 to 69 dB(A)
??Alarm system delays: Less than 10 seconds after detection (low oxygen alarms if oxygen is less than 82% volume fraction at specified environmental conditions)
Oxygen concentrator status indicator:
??High priority alarm that indicates when oxygen concentration drops below 82%
*At pulse setting 2.